ASTM D4282 Standard Test Method for Determination of Free Cyanide in Water and Wastewater by Microdiffusion
Данный раздел/документ содержится в продуктах:
- Техэксперт: Машиностроительный комплекс
- Картотека зарубежных и международных стандартов
- ASTM D2036 Standard Test Methods for Cyanides in Water
- 13
- ASTM D2105 Standard Test Method for Longitudinal Tensile Properties of “Fiberglass” (Glass-Fiber- Reinforced Thermosetting-Resin) Pipe and Tube
- ASTM E4 Standard Practices for Force Verification of Testing Machines
- ASTM E855 Standard Test Methods for Bend Testing of Metallic Flat Materials for Spring Applications Involving Static Loading
- ASTM E4 Standard Practices for Force Verification of Testing Machines
- ASTM E4 Standard Practices for Force Verification of Testing Machines
- ASTM E855 Standard Test Methods for Bend Testing of Metallic Flat Materials for Spring Applications Involving Static Loading
- ASTM E855 Standard Test Methods for Bend Testing of Metallic Flat Materials for Spring Applications Involving Static Loading
- ASTM E691 Standard Practice for Conducting an Interlaboratory Study to Determine the Precision of a Test Method
- ASTM E2242 Standard Test Method for Column Percolation Extraction of Mine Rock by the Meteoric Water Mobility Procedure
- ASTM E877 Standard Practice for Sampling and Sample Preparation of Iron Ores and Related Materials for Determination of Chemical Composition and Physical Properties
- ASTM D7572 Standard Guide for Recovery of Aqueous Cyanides by Extraction from Mine Rock and Soil
- Картотека зарубежных и международных стандартов
ASTM International
Standard Test Method for Determination of Free Cyanide in Water and Wastewater by Microdiffusion
N D4282
Annotation
This test method covers the determination of free cyanides in waters and wastewaters. Free cyanide is here defined as the cyanide which diffuses as cyanide (HCN), at room temperature, from a solution at pH 6.2
This test method does not include complexes that resist dissociation, such as hexacyanoferrates and gold cyanide, nor does it include thiocyanate and cyanohydrin.
This test method may be applied to water and wastewater samples containing free cyanide from 10 to 150 g/L. Greater concentrations may be determined by appropriate dilution.
This test method has been fully validated by collaborative testing as specified by Practice D2777.
The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. For specific hazard statements, see 8.6, 8.9, Section 9, and 12.2.1.
2 The paper by J. M. Kruse and L. E. Thibault "Determination of Free Cyanide in Ferro- and Ferricyanides," Analytical Chemistry, 45(13): 2260–2261; 1973 Nov., recommends a diffusion at pH 7. The ANSI modification (ANSI PH 4.41-1978) uses pH 6. Using the conditions of the ANSI method, diffusion is completed within 4 hours at pH 6. Longer diffusion time was required at pH 7 on the samples analyzed.