ASTM F2182 REV A Standard Test Method for Measurement of Radio Frequency Induced Heating On or Near Passive Implants During Magnetic Resonance Imaging
Данный раздел/документ содержится в продуктах:
- Техэксперт: Машиностроительный комплекс
- Картотека зарубежных и международных стандартов
- ISO 25539-2 Cardiovascular implants — Endovascular devices — Part 2: Vascular stents - First Edition
- 11
- ISO 25539-2 Cardiovascular implants — Endovascular devices — Part 2: Vascular stents - First Edition
- 11.040
- ISO 25539-2 Cardiovascular implants — Endovascular devices — Part 2: Vascular stents - First Edition
- 11.040.40
- IEC 60601-2-33 CORR 1 Medical electrical equipment – Part 2-33: Particular requirements for the basic safety and essential performance of magnetic resonance equipment for medical diagnosis - Edition 3.0
- IEC 60601-2-33 CORR 1 Medical electrical equipment – Part 2-33: Particular requirements for the basic safety and essential performance of magnetic resonance equipment for medical diagnosis - Edition 3.0
- IEC 60601-2-33 CORR 1 Medical electrical equipment – Part 2-33: Particular requirements for the basic safety and essential performance of magnetic resonance equipment for medical diagnosis - Edition 3.0
- ASTM F2182 REV A Standard Test Method for Measurement of Radio Frequency Induced Heating On or Near Passive Implants During Magnetic Resonance Imaging
- ASTM F2052 Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment
- Картотека зарубежных и международных стандартов
ASTM International
Standard Test Method for Measurement of Radio Frequency Induced Heating On or Near Passive Implants During Magnetic Resonance Imaging
N F2182 REV A
Annotation
This test method covers measurement of radio frequency (RF) induced heating on or near a passive medical implant and its surroundings during magnetic resonance imaging (MRI).
This test method is one required to determine if the presence of a passive implant may cause injury to the patient with the implant during an MR procedure. Other safety issues that should be addressed include magnetically induced displacement force and torque, as well as proper device function while in various configurations in the MR environment.
The amount of RF-induced temperature rise for a given specific absorption rate (SAR) will depend on the RF frequency, which is dependent on the static magnetic field strength of the MR system. While the focus in this test method is on 1.5 Tesla (T) or 3 Tesla cylindrical bore MR systems, the RF-induced temperature rise for an implant in MR systems of other static magnetic field strengths or magnet designs can be evaluated by suitable modification of the method described herein.
This test method assumes that testing is done on devices that will be entirely inside the body. For other implantation conditions (for example, external fixation devices, percutaneous needles, catheters or tethered devices such as ablation probes), modifications of this test method are necessary.
This test method applies to whole body magnetic resonance equipment, as defined in section 2.2.103 of the IEC Standard 60601-2-33, Ed. 2.0, with a whole body RF transmit coil as defined in section 2.2.100. The RF coil is assumed to have quadrature excitation.
The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.