ASTM F2888 Standard Test Method for Platelet Leukocyte Count-An In-Vitro Measure for Hemocompatibility Assessment of Cardiovascular Materials
Данный раздел/документ содержится в продуктах:
- Техэксперт: Машиностроительный комплекс
- Картотека зарубежных и международных стандартов
- ISO FDIS 10993-4 Biological evaluation of medical devices - Part 4: Selection of tests for interactions with blood
- 11
- ISO FDIS 10993-4 Biological evaluation of medical devices - Part 4: Selection of tests for interactions with blood
- 11.100
- ASTM F2888 Standard Test Method for Platelet Leukocyte Count-An In-Vitro Measure for Hemocompatibility Assessment of Cardiovascular Materials
- ASTM F2902 Standard Guide for Assessment of Absorbable Polymeric Implants
- ASTM F2579 Standard Specification for Amorphous Poly(lactide) and Poly(lactide-co-glycolide) Resins for Surgical Implants
- ASTM F748 Standard Practice for Selecting Generic Biological Test Methods for Materials and Devices
- ASTM F2382 Standard Test Method for Assessment of Intravascular Medical Device Materials on Partial Thromboplastin Time (PTT)
- Картотека зарубежных и международных стандартов
ASTM International
Standard Test Method for Platelet Leukocyte Count-An In-Vitro Measure for Hemocompatibility Assessment of Cardiovascular Materials
N F2888
Annotation
This test method assists in the evaluation of cardiovascular device materials for their ability to induce thrombus formation. Thrombus formation is assessed by means of a reduction in human platelets and leukocytes when consumed by thrombus after activation on the material surface. This assay may be part of the hemocompatibility evaluation for devices and materials contacting human blood, as per ANSI/AAMI/ ISO 10993–4. See also Test Method F2382.
All safety policies and practices shall be observed during the performance of this test method. All human blood and any materials that had contact with human blood shall be bagged in a biohazard bag, properly labeled as the contents, and disposed of by appropriate means.
The human blood should be handled at Biosafety Level 2 as recommended in the Centers for Disease Control/National Institutes of Health Manual Biosafety in Microbiological Laboratories. The human blood donor must have tested negative for Hepatitis B (HBV) and Human Immunodeficiency (HIV) viruses. The blood should be treated like any patient blood in using universal precautions.
The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. Some specific hazards statements are given in Section 7 on Hazards.