DS DS/ISO 16063-17 Methods for the calibration of vibration and shock transducers – Part 17: Primary calibration by centrifuge
DANSK - Dansk Standard
Methods for the calibration of vibration and shock transducers – Part 17: Primary calibration by centrifuge
N DS DS/ISO 16063-17
Annotation
ISO 16063 comprises a series of documents dealing with methods for the calibration of vibration and shock transducers. ISO 16063-17:2016 lays down detailed specifications for the instrumentation and procedure to be used for primary calibration of accelerometers using centrifuge calibration. ISO 16063-17:2016 is applicable to rectilinear accelerometers with zero-frequency response, mainly of the strain gauge or piezoresistive type, and to primary standard and working transducers. It is applicable for a calibration range from 10 m/s2 to 20 000 m/s2 (higher accelerations possible) at 0 Hz. The limits of uncertainty applicable are 1 % of reading.
Автоматический перевод:
Методы для калибровки вибрации и преобразователей шока – Часть 17: Основная калибровка центрифугой
ISO 16063 включает ряд документов, имеющих дело с методами для калибровки преобразователей шока и вибрации. 16063-17:2016 ISO устанавливает подробные спецификации для инструментария и процедуру, которая будет использоваться для основной калибровки акселерометров с помощью калибровки центрифуги. 16063-17:2016 ISO применим к прямолинейным акселерометрам с нулевой частотной характеристикой, в основном тензодатчика или пьезорезистивного типа, и к основному стандарту и рабочим преобразователям. Это применимо для калибровочного диапазона от 10 m/s2 до 20 000 m/s2 (более высокие возможные ускорения) на уровне 0 Гц. Пределы применимой неуверенности являются 1% чтения.